Innovative Method of Flow Profile Formation for Ultrasonic Flowmeters

Authors

DOI:

https://doi.org/10.5755/j02.eie.9126

Abstract

The ultrasonic flow sensor with minimized temperature-dependent hydrodynamic flow measurement errors is presented. The experimental investigation of ultrasonic flowmeter SDU-1L (manufacturer AB “Axis Industries”, Lithuania) with the flow sensor size DN20 and DN50 has been performed. The results of the investigation show that by inserting a flow profile forming equipment into a flow sensor measurement channel it is possible to minimize the slope of the temperature-dependent hydrodynamic correction factor more than 3 times. Such method of error minimization allows us to avoid the necessity to apply temperature corrections of hydrodynamic factors as well as to expand dynamic measurement range.

Downloads

Published

2010-11-19

Issue

Section

HIGH FREQUENCY TECHNOLOGY, MICROWAVES

How to Cite

Borodičas, P., Ragauskas, A., Petkus, V., Bagdonas, A., & Šliteris, R. (2010). Innovative Method of Flow Profile Formation for Ultrasonic Flowmeters. Elektronika Ir Elektrotechnika, 106(10), 91-94. https://doi.org/10.5755/j02.eie.9126